KANGJIAN ESR Fast Detector

KANGJIAN ESR Fast Detector

Score
Sold: 0 Question And Answer 0

MOQ: 100 Piece

Note: If you have any questions,please contact us

  • Better compatibility compared with Westergren method.
  • A safe, reliable and airtight operation during the whole process from blood collection to test, free of biological contamination.
  • Designed with spirit level of patent technology, which ensures the detector horizontal and aclinic.
  • 10 channels, synchronous operation allowed.
  • Only 30 minutes needed for reading, easy and fast.
  • Stainless steel soleplate with silicagel cushion on, which will avoid any noise or breakage when the ESR tube being placed.
  • 179-0013
Jiangsu KANGJIAN Medical Apparatus Co.,Ltd

Jiangsu KANGJIAN Medical Apparatus Co.,Ltd

Since:

1998

You May Also Like

Product Description

● How to Use:

1. Carry out the venous blood collection according to standard requirements with 9x120mm ESR vacuum tube.

2. Immediately invert the tube at 180 degrees for 6-8 times to achieve thorough mixing, which will avoid hemolysis, clotting or blood bubble.

3. At the room temperature of around 20℃,vertically place the ESR tube containing blood sample onto the detector, note down the starting

time and relevant numbers. Keep the detector still for 30 minutes and then read the millimeter of erythrocyte sedimentation.

4. Detailed reading method: Keep it stable for 30min, align the plasma concave in ESR tubes to the zero scare of the detector. Then, read the

data by aligning the upper surface of erythrocyte to the scale on the detector. (See as the schematic diagram)


● Notice:

1. Before operation, the ESR fast detector should be put on a steady platform and adjusted to be horizontal with knobs below the soleplate.

2. The ESR tubes should be kept vertical during the whole sedimentation process, angle of inclination should not be more than 3°. Orelse, the

testing result will be of large deviation.

3. Keep the room temperature at 20±5℃, sedimentation time 30±2min.